【适用范围】
Suitable for testing the thermal resistance and moisture resistance of flat materials such as textiles, garments, bedding, sleeping bags, indoor decorations, films, coated materials, foams, leathers, and textile combinations of multi-layer fabrics.
【相关标准】
【Product Features】
1.The detection accuracy reaches the world's advanced level, making it an ideal instrument for measuring the thermal and moisture performance indicators of textiles in terms of physiological comfort.
2.The core parts of the instrument adopt patented technology and high-precision temperature sensors, ensuring the testing accuracy of the instrument, as well as the perfect repeatability and stability of data.
3.Equipped with a microcomputer control and data processor, it can directly measure and calculate various properties of the sample (thermal resistance, moisture resistance, moisture permeability rate, moisture permeability index, heat retention rate, heat transfer coefficient, clo value) without manual calculation.
4.Adopts a large-screen color LCD touch screen display with a perfect UI design. It can store all data during the testing process and final statistical results for query.
5.Through API software, direct computer control and operation can be realized. It can display various temperature and humidity balance curves in real time, store all data during the testing process and final statistical results for real-time query, and output test results directly through the computer and printer.
6.Displays thermal resistance and moisture resistance data in real time, and performs multi-point independent calibration to improve the non-linearity of the system.
7.Uses an imported motor-driven lifting system to quickly adapt to the testing of samples with different thicknesses.
8.Equipped with a fully automatic water level replenishment system with an error of less than 1mm.
9.The internal water tank adopts anti-corrosion treatment technology with aviation materials, ensuring the durability and sturdiness of the precision parts of the instrument and greatly improving the overall stability of the instrument.
10.The ventilation system uses components of the German imported EBM brand, further improving the stability and reliability of internal air flow transmission.
11.Provides standard samples for testing the thermal resistance and moisture resistance of textiles.
【Technical Parameters】
1.Testing Methods:
Method A: Evaporative Hot Plate Method;
Method B: Static Flat Plate Method;
Testing of heat absorption and heat release of phase-change fabrics (a breakthrough technology, an international initiative).
2.Test Plate Size: 200×200×5mm;
The heating test plate assembly consists of three parts: the test plate, the thermal guard ring and the base plate. The thermal guard ring is located around the test plate, and the base plate is located below the test plate.
The test plate is a microporous porous plate made of special stainless steel powder metallurgy, which highly simulates the sweating effect of human skin. At 20℃, when a light beam with a wavelength range of 8μm to 14μm is vertically irradiated on the surface of the test plate and measured in a hemispherical reflection mode, the emissivity of the test plate surface is higher than 0.4.
3.Thermal Guard Ring Width: 125mm;
4.Temperature Adjustment Range of Test Plate, Thermal Guard Ring and Base Plate: 20~50℃ (adjustable);
Temperature Control Accuracy of Test Plate, Thermal Guard Ring and Base Plate: ±0.1℃; Resolution: 0.01℃;
5.Air Velocity Above the Sample Surface: 0.5, 1, 1.5 m/s (optional); ±0.05m/s; (The air velocity adjustment range is wider, with a broader scope of application, meeting the needs of various researches.)
6.Thermal Resistance:
(1) Testing Range: 0~2.0 m²•K/W;
(2) Resolution: 0.00001 m²•K/W;
(3) Repeatability Error: ≤±2.5%, factory control ±2%;
7.Moisture Resistance:
(1) Testing Range: 0~1000 m²•Pa/W;
(2) Resolution: 0.01 m²•Pa/W;
(3) Repeatability Error: ≤±2.5%, factory control ±2%;
8.Temperature Control Accuracy of Test Plate, Thermal Guard Ring and Base Plate: ±0.1℃; Resolution: 0.01℃;
9.Microclimate Room Temperature and Humidity/Accuracy:
For thermal resistance: 20℃±0.1℃ (Resolution: 0.01℃), 65%±2% (Resolution: 0.01%);
For moisture resistance: 35℃±0.1℃ (Resolution: 0.01℃), 40%±2% (Resolution: 0.01%);
10.Thermal Power Measurement Accuracy: ±0.5%.
11.Platform Lifting Range (Sample Thickness): 70mm;
12.Overall Dimensions: 1000×770×1800mm;
13.Weight: 400Kg;
14.Power Supply/Power: AC220V±10%/4000W.
15.Equipped with an automatic water supply system and a 15L large-capacity water storage bucket.
16.The instrument is configured with an upgraded API software, a computer and a printer.